Absorbance of the sample solution is 0.153AU. The three main uncertainty sources of absorbance in photometric analysis are repeatability, drift of the photometer and uncertainty due to possible interfering effects. The respective uncertainty components were estimated by the analyst as follows: Repeatability standard deviation is 0.004AU, uncertainty due to possible drift is ± 0.001AU and uncertainty due to the possible interfering effects is ± 2% of the absorbance value.

Answers

Answered by bannybannyavvari
3

Answer:

If we make a number of repeated measurements under the same conditions then the standard deviation of the obtained values characterized the uncertainty due to non-ideal repeatability (often called as repeatability standard uncertainty) of the measurement: u (V, REP) = s(V).