Chemistry, asked by krishnanaik7095, 1 year ago

Consider the following statements with the respect to the multi-channel peristaltic pump discussed in the course: S1: All the four channels in the multichannel peristaltic pump can circulate or flow liquid at the same flow rate only S2: The minimum flow rate possible in the channels is in the range of mL/min. Which of the above statements are correct?

Answers

Answered by sharjeel759
1

Answer:

None of the statement is correct.

Explanation:

S2: the minimum flow rate of perisitaltic pump is even below the mL/min, it's in the range of micro-liter/min like the range of mastertech FH100M multi channel peristaltic pump is between 1.2 miroliter/min to 760 mL/min. the maximum flow rate possible is in mL/min.

S1: A multichannel pump can have 4, 6, 8, 12 channels but flow rate in each channel is not equal to each other rather they are proportional to one another.

A peristaltic pump can have have proportionality of 100:1 of fluid flow between it's channels.

Hope the answer may be helpful!