Perform linearity testing of the timer & consistency of mA loading
Answers
Answered by
0
perform literal testing of the timer and the consisting of a loading
Answered by
0
To Perform linearity testing of the timer and consistency of mA loading:
- To place the dosimeter on the table on a sheet of lead rubber to absorb back-scatter.
- Ensure that the SID is 100cm to the top of the dosimeter.
- Take 4 exposures using 80 kV and 20 mAs. Each exposure should have different mA and time combinations equal to 20 mAs.
- Record the dosimeter readings from each exposure.
- Divide each reading by 20 mAs to get mRs/mAs.
- Calculate Linearity variance.